DESCRIPTION
Used for the quantitative determination of Anti-Infliximab concentration in serum and plasma.
PRINCIPLE OF THE ASSAY
This assay employs the quantitative sandwich enzyme immunoassay technique. Infliximab has been pre-coated onto a microplate. Samples or standards are pipetted into microwells and Anti-Infliximab will be captured by immobilized Infliximab. After washing away any unbound substances, a biotin-labeled Infliximab is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Anti-Infliximab bound in the initial step. The color development is stopped and the intensity of the color is measured.
PRECISION
Intra-Assay Precision (Precision within an assay): <10% Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision. Inter-Assay Precision (Precision between assays): <15% Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (ng/mL) 241.6 58.4 9.4 235.6 48.7 9.9 Standard deviation 8.1 1.8 0.9 14.2 1.8 0.6 CV \(%\) 3.4 3.0 9.9 6.0 3.7 6.4 RECOVERY
The recovery of Anti-Infliximab spiked to three different levels in human serum samples throughout the range of the assay was evaluated.
Sample Dilution factor Average (%) Range (%) Human serum (n=3) 1:10 92 88-97 1:20 92 86-97 LINEARITY
To assess the linearity of the assay, human serum samples were spiked with high concentrations of Anti-Infliximab and diluted with the Standard Diluent to produce samples with values within the dynamic range of the assay.
Sample Dilution factor Average (%) Range (%) Human serum (n=4) 1:2 101 98-102 1:4 103 101-105 1:8 104 100-107 1:16 94 80-101