DESCRIPTION
A glycosylphosphatidyl-inositol-(GPI)-anchored glycoprotein, carcinoembryonic antigen (CEA)-related cell adhesion molecule 8 (CEACAM8), also known as CD66b, NCA-95 and CD67, was identified in human neutrophils. It is solely expressed by granulocytes, of which 95% are polymorphonuclear neutrophils (PMN). CEACAM8 is cell-adhesion proteins on neutrophils that belong to the human carcinoembryonic antigen (CEA) family. CEA family members are believed to function as intercellular adhesion molecules in vitro. CEACAM8 has been shown to adhere to the closely related molecule CEACAM6 (NCA-50/90 and CD66c), suggesting a role in the interaction between granulocytes or between granulocytes and epithelial cells expressing CEACAM6 in vivo.
PRINCIPLE OF THE ASSAY
Sandwich ELISA structure (Detection antibody labeled with HRP) A capture antibody is pre-coated onto the bottom of wells which binds to analyte of interest. A detection antibody labeled with HRP also binds to the analyte. TMB acts as the HRP substrate and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns solution yellow. The color intensity is proportional to the quantity of bound protein which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 20 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 24 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 20 3,048.4 99.5 3.3 2 20 743.6 23.1 3.1 3 20 191.2 13.3 7.0 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 24 3,055.9 101.5 3.3 2 24 709.0 18.7 2.6 3 24 180.5 8.5 4.7 RECOVERY
The recovery of human CEACAM8/CD66b spiked to three different levels throughout the range of the assay in various matrices were evaluated.
Sample Type Average% of Expected Range (%) Human serum 1:2 90 80-101 1:4 104 92-119 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay.
Human serum 1:2 Average% of Expected 100 Range (%) - 1:4 Average% of Expected 110 Range (%) 101-117 1:8 Average% of Expected 102 Range (%) 93-114 1:16 Average% of Expected 100 Range (%) 93-111 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human Human CEACAM8/CD66b Sandwich ELISA Kit produced at Sunlong Medical™.