DESCRIPTION
Neutrophil elastase (NE) is a major proteinase in primary granules in neutrophils that participates in microbicidal activity. NE (Leukocyte elastase, EC 3.4.21.37), a serine endopeptidase, is characterized by serine in the active tripeptide catalytic site: Asp, His, Ser. NE is the most abundant of four serine proteases present in neutrophils, which also include proteinase 3, cathepsin G, and neutrophil serine protease 4 (NSP4). NE is a 29.5 kD protein stored in mature form in the azurophilic granules of neutrophils and is present at high concentrations per azurophilic granule-approximately 67,000 molecules per granule.
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 1,249.0 47.7 3.8 2 8 264.5 8.8 3.3 3 8 139.5 4.1 2.9 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 1,228.0 60.3 4.9 2 16 242.7 24.5 10.1 3 16 128.5 12.8 10.0 RECOVERY
The recovery of mouse Neutrophil Elastase spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Mouse serum 1:32 95 87-109 Tissue lysate 1:4 103 83-117 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The mouse serum was initially diluted 1:2)
Mouse serum(Sample Diluent PT 1-ef) Tissue lysate(Sample Diluent PT 5-ef) 1:2 Average% of Expected 100 100 Range (%) - - 1:4 Average% of Expected 106 111 Range (%) 105-106 98-124 1:8 Average% of Expected 111 101 Range (%) 108-114 89-113 1:16 Average% of Expected 100 88 Range (%) 99-102 77-99 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant mouse Neutrophil Elastase produced at Sunlong Medical™.