DESCRIPTION
Used for the quantitative determination of Anti-Atezolizumab concentration in serum and plasma.
PRINCIPLE OF THE ASSAY
This assay employs the quantitative sandwich enzyme immunoassay technique. Atezolizumab has been pre-coated onto a microplate. Samples or standards are pipetted into microwells and Anti-Atezolizumab will be captured by immobilized Atezolizumab. After washing away any unbound substances, a biotin-labeled Atezolizumab is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Anti-Atezolizumab bound in the initial step. The color development is stopped and the intensity of the color is measured.
PRECISION
Intra-Assay Precision (Precision within an assay): <10% Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision. Inter-Assay Precision (Precision between assays): <15% Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (ng/mL) 47.6 11.8 2.5 49.0 11.7 3.1 Standard deviation 1.3 0.5 0.1 1.8 0.4 0.2 CV \(%\) 2.8 4.4 5.6 3.7 3.6 7.0 RECOVERY
The recovery of Anti-Atezolizumab spiked to three different levels in human serum samples throughout the range of the assay was evaluated.
Sample Dilution factor Average (%) Range (%) Human serum (n=4) 1:10 93 82-108 1:20 93 83-103 LINEARITY
To assess the linearity of the assay, human serum samples were spiked with high concentrations of Anti-Atezolizumab and diluted with the Standard Diluent to produce samples with values within the dynamic range of the assay.
Sample Dilution factor Average (%) Range (%) Human serum (n=4) 1:2 96 93-98 1:4 96 93-100 1:8 99 83-115 1:16 99 87-110