DESCRIPTION
Used for the quantitative determination of Anti-Rituximab concentration in serum and plasma.
PRINCIPLE OF THE ASSAY
This assay employs the quantitative sandwich enzyme immunoassay technique. Rituximab has been pre-coated onto a microplate. Samples or standards are pipetted into microwells and Anti-Rituximab will be captured by immobilized Rituximab. After washing away any unbound substances, a biotin-labeled Rituximab is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Anti-Rituximab bound in the initial step. The color development is stopped and the intensity of the color is measured.
PRECISION
Intra-Assay Precision (Precision within an assay): <10% Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision. Inter-Assay Precision (Precision between assays): <15% Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (ng/mL) 32.1 8.6 2.1 32.2 8.8 2.4 Standard deviation 0.8 0.2 0.1 1.7 0.3 0.2 CV \(%\) 2.5 2.3 4.2 5.1 3.0 8.1 RECOVERY
The recovery of Anti-Rituximab spiked to three different levels in human serum samples throughout the range of the assay was evaluated.
Sample Dilution factor Average (%) Range (%) Human serum (n=4) 1:10 91 80-112 1:20 89 85-95 LINEARITY
To assess the linearity of the assay, human serum samples were spiked with high concentrations of Anti-Rituximab and diluted with the Standard Diluent to produce samples with values within the dynamic range of the assay.
Sample Dilution factor Average (%) Range (%) Human serum (n=4) 1:2 96 91-99 1:4 107 104-111 1:8 118 114-120 1:16 106 100-113