DESCRIPTION
Used for the quantitative determination of Otilimab concentration in serum and plasma.
PRINCIPLE OF THE ASSAY
This assay employs the quantitative competitive enzyme immunoassay technique. Recombinant Human CSF2 has been pre-coated onto a microplate. Standards or samples are premixed with biotin-labeled antibody and then pipetted into the wells. Otilimab in the sample competitively binds to the pre-coated protein with biotin-labeled Otilimab. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in inversely proportion to the amount of Otilimab bound in the initial step. The color development is stopped and the intensity of the color is measured.
PRECISION
Intra-Assay Precision (Precision within an assay): <20% Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision. Inter-Assay Precision (Precision between assays): <20% Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (ng/mL) 4393.1 1025.1 323.1 3625.9 871.8 312.0 Standard deviation 248.5 117.1 56.2 234.9 73.4 48.0 CV \(%\) 5.7 11.4 17.4 6.5 8.4 15.4 RECOVERY
The recovery of Otilimab spiked to three different levels in human serum samples throughout the range of the assay was evaluated.
Sample Dilution factor Average (%) Range (%) Human serum (n=3) 1:10 88 80-98 1:20 83 80-89 LINEARITY
To assess the linearity of the assay, human serum samples were spiked with high concentrations of Otilimab and diluted with the Standard Diluent to produce samples with values within the dynamic range of the assay.
Sample Dilution factor Average (%) Range (%) Human serum (n=4) 1:2 94 93-95 1:4 104 94-109 1:8 105 98-111