Neutrophil elastase (NE) is a major proteinase in primary granules in neutrophils that participates in microbicidal activity. NE (Leukocyte elastase, EC 3.4.21.37), a serine endopeptidase, is characterized by serine in the active tripeptide catalytic site: Asp, His, Ser. NE is the most abundant of four serine proteases present in neutrophils, which also include proteinase 3, cathepsin G, and neutrophil serine protease 4 (NSP4). NE is a 29.5 kD protein stored in mature form in the azurophilic granules of neutrophils and is present at high concentrations per azurophilic granule-approximately 67,000 molecules (~5 mM) per granule.
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 1,102.5 32.3 2.9 2 8 243.1 5.8 2.4 3 8 124.2 2.9 2.3 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 1,100.3 55.6 5.1 2 16 243.9 6.2 2.5 3 16 126.9 5.0 3.9 RECOVERY
The recovery of human Neutrophil Elastase spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human plasma 1:320 104 96-115 1:640 103 98-108 Saliva 1:1,600 105 101-107 1:3,200 110 103-121 Cell lysate 1:64,000 94 91-98 1:128,000 94 91-99 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The human plasma was initially diluted 1:40. The saliva was initially diluted 1:200. The cell lysate was initially diluted 1:8,000.)
Human plasma Saliva Cell lysate 1:2 Average% of Expected 100 100 100 Range (%) - - - 1:4 Average% of Expected 101 99 94 Range (%) 100-102 97-100 93-96 1:8 Average% of Expected 97 93 91 Range (%) 97-98 90-97 91-92 1:16 Average% of Expected 98 92 85 Range (%) 94-102 91-93 83-87 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human Neutrophil Elastase produced at Sunlong Medical™.