MPO(myeloperoxidase) is a peroxidase enzyme presented in the azurophilic granules of polymorphonuclear (PMN) leukocytes and monocytes. Plasma concentration of MPO can be used as a specific marker of PMN activation. MPO catalyzes the production of hypochlorous acid (HClO) from hydrogen peroxide (H2O2) and chloride anion (Cl-, or the equivalent from a non-chlorine halide). This enzymatic system plays an important role in human defense against microorganisms. The serum/plasma MPO levels have been associated with a variety of clinical conditions including inflammatory diseases, atherosclerosis, ischaemic stroke, hypertension, heart failure, risk of cardiovascular events and so on.
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 1,242.7 31.5 2.5 2 8 292.6 7.5 2.6 3 8 143.4 5.1 3.5 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 1,259.0 34.6 2.8 2 16 307.2 16.8 5.5 3 16 149.2 8.5 5.7 RECOVERY
The recovery of human MPO spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human serum 1:3,200 108 101-121 Cell culture supernatant 1:2,560 105 91-118 Urine 1:16 109 108-110 Saliva 1:6,400 90 88-95 Cell lysate 1:16 110 100-127 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The human serum was initially diluted 1:200. The cell culture supernatant was initially diluted 1:160. The saliva was initially diluted 1:400. The cell lysate was initially diluted 1:10,000.)
Human serum Cell culturesupernatant Urine Saliva Cell lysate 1:2 Average% ofExpected 100 100 100 100 100 Range (%) - - - - - 1:4 Average% ofExpected 88 85 91 88 95 Range (%) 81-95 82-87 89-94 78-98 84-105 1:8 Average% ofExpected 120 93 92 82 81 Range (%) 111-130 88-97 77-108 77-87 80-82 1:16 Average% ofExpected 106 97 102 84 89 Range (%) 95-116 91-103 88-116 70-98 79-99 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human MPO produced at Sunlong Medical™.