TEL: +86 571 56623320 EMAIL: [email protected]
DESCRIPTION
Lactoferrin (LTF), known as an iron-binding glycoprotein, is the most abundant element in milk. Prior research has illustrated that LTFbears multi-pharmacological properties, including protection against infection, regulation inflammatory response, antioxidant and antifibrotic process. Moreover, clinical trials have identified the safety of oral administration of bovine LTF and demonstrated the efficacy of LTF on anti-infection, treatment of inflammation, cancer, metabolic disorders.
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 537.2 22.8 4.2 2 8 140.3 14.7 10.5 3 8 71.0 2.05 2.9 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 570.2 40.0 7.0 2 16 141.1 13.2 9.4 3 16 70.9 2.4 3.3 RECOVERY
The recovery of human Lactoferrin/LTF spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human plasma 1:800 99 95-102 1:1,600 84 76-93 Urine 1:32 96 90-102 Saliva 1:51,200 92 85-98 1:102,400 98 89-109 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The human plasma was initially diluted 1:100. The urine was initially diluted 1:8. The saliva samples were initially diluted 1:6,400.)
Human plasma Urine Saliva 1:2 Average% of Expected 100 100 100 Range (%) - - - 1:4 Average% of Expected 95 104 104 Range (%) 90-100 94-114 99-108 1:8 Average% of Expected 105 108 101 Range (%) 99-111 101-115 100-102 1:16 Average% of Expected 104 103 106 Range (%) 97-111 96-110 105-107 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human Human Lactoferrin/LTF One-Step ELISA Kit produced at Sunlong Medical™.