DESCRIPTION
Thrombomodulin (THBD, also known as CD141) is an endothelial-specific type I membrane glycoprotein of 105 kDa that binds thrombin, which plays an important role as a natural anticoagulant. It is expressed mostly on the surface of vascular endothelial cells, where it probably be involved in cell differentiation or inflammation. Soluble forms of CD141 have been found in plasma and urine. It has been reported that plasma CD141 is increased in patients with diabetes mellitus, particularly those with diabetic nephropathy.
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 4,059.3 221.3 5.5 2 8 1,041.6 60.1 5.8 3 8 306.5 8.8 2.9 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 4,014.1 168.0 4.2 2 16 1,005.6 59.2 5.9 3 16 302.2 19.3 6.4 RECOVERY
The recovery of human CD141/Thrombomodulin spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human plasma 1:8 104 101-106 1:16 114 103-131 Urine 1:16 94 84-99 1:32 96 90-100 Cell lysate 1:16 96 93-98 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The urine was initially diluted 1:4)
Human plasma(Sample Diluent PT 3-ef) Urine(Sample Diluent PT 4B1) Cell lysate(Sample Diluent PT 3-ef) 1:2 Average% of Expected 100 100 100 Range (%) - - - 1:4 Average% of Expected 119 87 103 Range (%) 116-122 81-93 102-103 1:8 Average% of Expected 101 89 110 Range (%) 98-104 85-93 110-111 1:16 Average% of Expected 96 90 111 Range (%) 93-100 85-94 108-114 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human Human CD141/Thrombomodulin One-Step ELISA Kit produced at Sunlong Medical™.