DESCRIPTION
Ubiquitin C-terminal hydrolase 1 (UCHL1), previously known as PGP9.5, is found primarily in neurons and neuroendocrine cells with a very high concentration in dendrites and perikarya. UCHL1 is a more recently recognized biomarker of neuronal injury. UCHL1 could be detected in the CSF and serum with a half-life of 7-9 hours after severe traumatic brain injury. Increased plasma concentrations of UCHL1 therefore reflect increased blood brain barrier permeability and destruction of neurons.
PRINCIPLE OF THE ASSAY
Sandwich ELISA structure (Detection antibody labeled with HRP) A capture antibody is pre-coated onto the bottom of wells which binds to analyte of interest. A detection antibody labeled with HRP also binds to the analyte. TMB acts as the HRP substrate and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns solution yellow. The color intensity is proportional to the quantity of bound protein which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 20 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 24 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (ng/mL) SD CV% 1 20 4.58 0.16 3.4 2 20 1.14 0.03 2.9 3 20 0.30 0.02 5.9 Inter-assay Precision
Sample n Mean (ng/mL) SD CV% 1 24 4.54 0.22 4.9 2 24 1.02 0.03 3.1 3 24 0.23 0.02 8.6 RECOVERY
The recovery of human UCH-L1 spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human serum 1:8 89 84-98 1:16 95 91-99 Cell lysate 1:800 108 92-123 1:1600 99 72-127 LINEARITY
To assess the linearity of the assay, human serum, cell lysates were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The serum samples were initially diluted 1:2. The cell lysate samples were initially diluted 1:200. )
Human serum(Sample Diluent PT 1B5) Cell lysate(Sample Diluent PT 5) 1:2 Average% of Expected 100 100 Range (%) - - 1:4 Average% of Expected 81 90 Range (%) 78-89 76-103 1:8 Average% of Expected 102 80 Range (%) 96-110 74-90 1:16 Average% of Expected 102 82 Range (%) 100-105 72-98 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human UCH-L1 produced at Sunlong Medical™.
Ordering & shipping
- Minimum order: 1 kit
- Lead time: varies with order quantity and specifications. Please consult your local distributor or our sales team for confirmation
- Payment: T/T 100% in advance; PayPal; Alibaba Trade Assurance
- Shipping: FedEx / DHL / UPS express; cold-chain packing available on request
- Price: USD, per kit (48T / 96T options available)
+86 571 56623320
[email protected]
SUNLONG BIOTECH

Scan Wechat Qrcode