TEL: +86 571 56623320 EMAIL: [email protected]
DESCRIPTION
Prolactin (PRL) is a pleiotropic hormone which is mainly synthetized by the lactotrophs of the anterior pituitary gland. Prolactin has a wide variety of effects. It causes the breasts to grow and make milk during pregnancy and after birth. Prolactin levels are normally high for pregnant women and new mothers. Prolactin also plays an essential role in metabolism, regulation of the immune system and pancreatic development. Normally during pregnancy prolactin serum levels are increased to hold of insulin resistance (IR). As well, prolactin is involved in regulation of glucose metabolism, since high prolactin levels are associated with low incidence of type 2 diabetes mellitus (T2DM).
PRINCIPLE OF THE ASSAY
Sandwich ELISA structure (Detection antibody labeled with biotin) A capture antibody is pre-coated onto the bottom of wells which binds to analyte of interest. A detection antibody labeled with biotin also binds to the analyte. Streptavidin-HRP binds to the biotin. TMB acts as the HRP substrate and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns solution yellow. The color intensity is proportional to the quantity of bound protein which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 20 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 24 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 20 101.7 8.5 8.4 2 20 305.6 5.6 1.8 3 20 1,054.2 21.8 2.1 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 24 72.9 4.1 5.1 2 24 264.1 7.1 2.7 3 24 958.0 28.2 2.9 RECOVERY
The recovery of human Prolactin spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human serum 1:8 94 89-100 1:16 103 102-110 Cell culture supernatant 1:2 101 95-109 1:4 85 78-90 Human milk 1:60 102 94-107 1:120 103 89-123 LINEARITY
To assess the linearity of the assay, cell culture supernatant samples were spiked with high concentrations of human Prolactin in various matrices and diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. Human serum and human milk were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay.(The human milk samples were initially diluted 1: 8 )
Human serum(Sample Diluent PT 1-eg) Cell culture supernatant(Sample Diluent PT 1-ag) Human milk(Sample Diluent PT 3) 1:2 Average% of Expected 100 113 100 Range (%) - 96-124 - 1:4 Average% of Expected 100 106 104 Range (%) 97-103 96-116 103-106 1:8 Average% of Expected 101 100 108 Range (%) 97-107 91-112 107-111 1:16 Average% of Expected 109 99 111 Range (%) 103-112 88-113 108-117 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human Prolactin produced at Sunlong Medical™.
Ordering & shipping
- Minimum order: 1 kit
- Lead time: varies with order quantity and specifications. Please consult your local distributor or our sales team for confirmation
- Payment: T/T 100% in advance; PayPal; Alibaba Trade Assurance
- Shipping: FedEx / DHL / UPS express; cold-chain packing available on request
- Price: USD, per kit (48T / 96T options available)
+86 571 56623320
[email protected]
SUNLONG BIOTECH

Scan Wechat Qrcode