TEL: +86 571 56623320 EMAIL: [email protected]
DESCRIPTION
Peroxin (PEX) proteins generate and maintain peroxisomes. The peroxin-1 (PEX1) ATPase facilitates the recycling of the peroxisome matrix protein receptor PEX5 and is the most commonly affected peroxin in human peroxisome biogenesis disorders. PEX1 and PEX6 are the only members of the AAA family (for ATPases associated with diverse cellular) activities implicated in peroxisome biogenesis and are closely related to p97, which functions in endoplasmic reticulum-associated protein degradation to retrotranslocate endoplasmic reticulum proteins to the cytosol.
PRINCIPLE OF THE ASSAY
Sandwich ELISA structure (Detection antibody labeled with HRP) A capture antibody is pre-coated onto the bottom of wells which binds to analyte of interest. A detection antibody labeled with HRP also binds to the analyte. TMB acts as the HRP substrate and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns solution yellow. The color intensity is proportional to the quantity of bound protein which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (ng/mL) SD CV% 1 8 50.85 3.07 6.04 2 8 13.34 1.16 8.73 3 8 5.98 0.30 4.96 Inter-assay Precision
Sample n Mean (ng/mL) SD CV% 1 16 54.55 4.64 8.51 2 16 13.27 1.33 10.03 3 16 5.52 0.55 9.87 RECOVERY
The recovery of human PEX1 spiked to three different levels throughout the range of the assay in cell lysate was evaluated.
Sample Type Average% of Expected Range (%) Cell lysate 1:16 84 74-92 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The cell lysate was initially diluted 1:2.)
Cell lysate 1:2 Average% of Expected 100 Range (%) - 1:4 Average% of Expected 123 Range (%) 118-129 1:8 Average% of Expected 120 Range (%) 113-127 1:16 Average% of Expected 103 Range (%) 92-115 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human Human PEX1 Sandwich ELISA Kit produced at Sunlong Medical™.
Ordering & shipping
- Minimum order: 1 kit
- Lead time: varies with order quantity and specifications. Please consult your local distributor or our sales team for confirmation
- Payment: T/T 100% in advance; PayPal; Alibaba Trade Assurance
- Shipping: FedEx / DHL / UPS express; cold-chain packing available on request
- Price: USD, per kit (48T / 96T options available)