DESCRIPTION
CD44, a transmembrane cell surface glycoprotein, is normally expressed in all human cells. It is primarily involved in regulating cell adhesion, migration, angiogenesis, proliferation, and inflammation. It may take part in the process of tumorigenesis as CD44 and its variants have a potential role in the development of malignant tumors. CD44 has gained significant attention because of its usefulness as a stem cell marker and has surfaced as a budding therapeutic target, imposing a greater understanding of CD44 in various cancers. Additionally, Serum CD44 concentration correlated with tumor burden and metastasis of tumors.
PRINCIPLE OF THE ASSAY
Sandwich ELISA structure (Detection antibody labeled with HRP) A capture antibody is pre-coated onto the bottom of wells which binds to analyte of interest. A detection antibody labeled with HRP also binds to the analyte. TMB acts as the HRP substrate and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns solution yellow. The color intensity is proportional to the quantity of bound protein which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 3,116.9 152.9 4.9 2 8 789.2 43.4 5.5 3 8 415.6 37.0 8.9 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 3,077.1 209.8 6.8 2 16 800.1 44.2 5.5 3 16 410.4 31.5 7.7 RECOVERY
The recovery of human CD44 spiked to three different levels throughout the range of the assay in human serum was evaluated.
Sample Type Average% of Expected Range (%) Human serum 1:400 101 98-106 1:800 113 109-115 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The human serum was initially diluted 1:50. )
Human serum 1:2 Average% of Expected 100 Range (%) - 1:4 Average% of Expected 100 Range (%) 96-109 1:8 Average% of Expected 94 Range (%) 84-104 1:16 Average% of Expected 96 Range (%) 85-107 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human CD44 produced at Sunlong Medical™.
Ordering & shipping
- Minimum order: 1 kit
- Lead time: varies with order quantity and specifications. Please consult your local distributor or our sales team for confirmation
- Payment: T/T 100% in advance; PayPal; Alibaba Trade Assurance
- Shipping: FedEx / DHL / UPS express; cold-chain packing available on request
- Price: USD, per kit (48T / 96T options available)
+86 571 56623320
[email protected]
SUNLONG BIOTECH

Scan Wechat Qrcode