Assay Principle
This kit employs double antibody sandwich ELISA technology: Capture Antibody is coated on the microplate to capture TH from samples and standards. After washing, biotin-labeled Detection Antibody is added and incubated, followed by washing to form a "Capture Antibody-
Antigen-Detection Antibody" immune-complex. Subsequently, Streptavidin-Horseradish Peroxidase (SA-HRP) is added and incubated. After incubation and washing, TMB Substrate is added for color development. If the target analyte is present in the sample, a blue color develops. Stop Solution is then added to terminate the reaction. During the assay, unbound components are washed away. The Optical Density (OD) is measured at 450 nm using a microplate reader. The intensity of the color is proportional to the TH concentration in the sample, and the concentration is calculated by plotting a standard curve.
Typical Data
The following data and curve are for reference only. The experimenter must establish a standard curve based on their own experimental data.
Standard Conc. (ng/mL)
20
10
5
2.5
1.25
0.625
0.313
0
2.706
1.531
0.792
0.425
0.222
0.134
0.085
0.04
Corrected OD Value
2.666
1.491
0.752
0.385
0.182
0.094
0.045
0
Sensitivity
The minimum detectable concentration (sensitivity) of Human TH, determined by testing samples, is 0.155ng/mL.