Used for the quantitative determination of Anti-Nivolumab concentration in serum and plasma.
PRINCIPLE OF THE ASSAY
This assay employs the quantitative sandwich enzyme immunoassay technique. Nivolumab has been pre-coated onto a microplate. Samples or standards are pipetted into microwells and Anti-Nivolumab will be captured by immobilized Nivolumab. After washing away any unbound substances, a biotin-labeled Nivolumab is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Anti-Nivolumab bound in the initial step. The color development is stopped and the intensity of the color is measured.
PRECISION
Intra-Assay Precision (Precision within an assay): <10% Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision. Inter-Assay Precision (Precision between assays): <15% Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (ng/mL) 101.1 25.5 6.8 110.2 27.1 7.6 Standard deviation 3.8 0.8 0.6 4.0 1.3 0.5 CV \(%\) 3.7 3.3 8.1 3.6 4.6 6.7 RECOVERY
The recovery of Anti-Nivolumab spiked to three different levels in human serum samples throughout the range of the assay was evaluated.
Sample Dilution factor Average (%) Range (%) Human serum (n=3) 1:10 92 81-111 1:20 87 80-92 LINEARITY
To assess the linearity of the assay, human serum samples were spiked with high concentrations of Anti-Nivolumab and diluted with the Standard Diluent to produce samples with values within the dynamic range of the assay.
Sample Dilution factor Average (%) Range (%) Human serum (n=3) 1:2 105 104-107 1:4 113 108-120 1:8 110 107-112 1:16 110 97-120