DESCRIPTION
Used for the quantitative determination of Anti-Daratumumab concentration in serum and plasma.
PRINCIPLE OF THE ASSAY
This assay employs the quantitative sandwich enzyme immunoassay technique. Daratumumab has been pre-coated onto a microplate. Samples or standards are pipetted into microwells and Anti-Daratumumab will be captured by immobilized Daratumumab. After washing away any unbound substances, a biotin-labeled Daratumumab is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Anti-Daratumumab bound in the initial step. The color development is stopped and the intensity of the color is measured.
PRECISION
Intra-Assay Precision (Precision within an assay): <10% Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision. Inter-Assay Precision (Precision between assays): <15% Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (ng/mL) 833.0 201.1 85.2 835.9 210.9 89.5 Standard deviation 33.1 11.1 8.8 37.7 16.9 10.3 CV \(%\) 4.0 5.5 10.4 4.5 8.0 11.5 RECOVERY
The recovery of Anti-Daratumumab spiked to three different levels in human serum samples throughout the range of the assay was evaluated.
Sample Dilution factor Average (%) Range (%) Human serum (n=3) 1:10 91 82-106 1:20 88 81-97 LINEARITY
To assess the linearity of the assay, human serum samples were spiked with high concentrations of Anti-Daratumumab and diluted with the Standard Diluent to produce samples with values within the dynamic range of the assay.
Sample Dilution factor Average (%) Range (%) Human serum (n=5) 1:2 98 92-105 1:4 102 93-112 1:8 109 100-116 1:16 104 95-113