DESCRIPTION
Used for the quantitative determination of Tocilizumab concentration in serum and plasma.
PRINCIPLE OF THE ASSAY
This assay employs the quantitative competitive enzyme immunoassay technique. Recombinant Human IL6R has been pre-coated onto a microplate. Standards or samples are premixed with biotin-labeled antibody and then pipetted into the wells. Tocilizumab in the sample competitively binds to the pre-coated protein with biotin-labeled Tocilizumab. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in inversely proportion to the amount of Tocilizumab bound in the initial step. The color development is stopped and the intensity of the color is measured.
PRECISION
Intra-Assay Precision (Precision within an assay): <20% Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision. Inter-Assay Precision (Precision between assays): <20% Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (ng/mL) 4802.5 1358.4 255.8 4927.2 1288.7 261.6 Standard deviation 367 115 26 557 170 31 CV \(%\) 7.7 8.5 10.0 11.3 13.2 11.9 RECOVERY
The recovery of Tocilizumab spiked to three different levels in human serum samples throughout the range of the assay was evaluated.
Sample Dilution factor Average (%) Range (%) Human serum (n=2) 1:10 103 88-117 1:20 115 94-126 LINEARITY
To assess the linearity of the assay, human serum samples were spiked with high concentrations of Tocilizumab and diluted with the Standard Diluent to produce samples with values within the dynamic range of the assay.
Sample Dilution factor Average (%) Range (%) Human serum (n=2) 1:2 102 89-115 1:4 100 97-102 1:8 98 93-104 1:16 101 101-102