DESCRIPTION
Used for the quantitative determination of Anti-Tocilizumab concentration in serum and plasma.
PRINCIPLE OF THE ASSAY
This assay employs the quantitative sandwich enzyme immunoassay technique. Tocilizumab has been pre-coated onto a microplate. Samples or standards are pipetted into microwells and Anti-Tocilizumab will be captured by immobilized Tocilizumab. After washing away any unbound substances, a biotin-labeled Tocilizumab is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Anti-Tocilizumab bound in the initial step. The color development is stopped and the intensity of the color is measured.
PRECISION
Intra-Assay Precision (Precision within an assay): <10% Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision. Inter-Assay Precision (Precision between assays): <15% Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (ng/mL) 46.0 10.9 2.1 41.2 11.9 2.7 Standard deviation 1.6 0.3 0.2 3.8 1.4 0.3 CV \(%\) 3.5 3.2 9.3 9.3 12.0 10.6 RECOVERY
The recovery of Anti-Tocilizumab spiked to three different levels in human serum samples throughout the range of the assay was evaluated.
Sample Dilution factor Average (%) Range (%) Human serum (n=3) 1:10 98 82-110 1:20 88 81-98 LINEARITY
To assess the linearity of the assay, human serum samples were spiked with high concentrations of Anti-Tocilizumab and diluted with the Standard Diluent to produce samples with values within the dynamic range of the assay.
Sample Dilution factor Average (%) Range (%) Human serum (n=3) 1:2 102 99-106 1:4 109 104-116 1:8 111 107-116 1:16 111 97-119