DESCRIPTION
Used for the quantitative determination of Anti-Trastuzumab concentration in serum and plasma.
PRINCIPLE OF THE ASSAY
This assay employs the quantitative sandwich enzyme immunoassay technique. Trastuzumab has been pre-coated onto a microplate. Samples or standards are pipetted into microwells and Anti-Trastuzumab will be captured by immobilized Trastuzumab. After washing away any unbound substances, a biotin-labeled Trastuzumab is added to the wells. After washing away any unbound substances, Streptavidin-HRP is added to the wells. Following a wash to remove any unbound enzyme reagent, a substrate solution is added to the wells and color develops in proportion to the amount of Anti-Trastuzumab bound in the initial step. The color development is stopped and the intensity of the color is measured.
PRECISION
Intra-Assay Precision (Precision within an assay): <10% Three samples of known concentration were tested sixteen times on one plate to assess intra-assay precision. Inter-Assay Precision (Precision between assays): <15% Three samples of known concentration were tested in twenty four separate assays to assess inter-assay precision.
Sample 1 2 3 1 2 3 n 16 16 16 24 24 24 Mean (ng/mL) 92.9 25.0 10.2 89.4 23.1 10.5 Standard deviation 4.2 1.4 0.8 8.8 1.9 0.9 CV \(%\) 4.5 5.6 7.9 9.9 8.3 8.2 RECOVERY
The recovery of Anti-Trastuzumab spiked to three different levels in human serum samples throughout the range of the assay was evaluated.
Sample Dilution factor Average (%) Range (%) Human serum (n=3) 1:10 100 94-111 1:20 99 90-106 LINEARITY
To assess the linearity of the assay, human serum samples were spiked with high concentrations of Anti-Trastuzumab and diluted with the Standard Diluent to produce samples with values within the dynamic range of the assay.
Sample Dilution factor Average (%) Range (%) Human serum (n=3) 1:2 98 94-101 1:4 104 100-111 1:8 101 94-113 1:16 95 83-101