Summary:
![]()
Purpose
Our Quick Step Fish malondialdehyde (MDA) ELISA Kit is to assay MDA levels in Fish serum, plasma, culture media or any biological fluid.
Principle
This ELISA kit uses Sandwich-ELISA as the method. The Microelisa stripplate provided in this kit has been pre-coated with an Antibody specific to MDA. Standards or samples are added to the appropriate Microelisa stripplate wells and combined to the specific Antibody. Then a Horseradish Peroxidase (HRP)- conjugated Antibody specific for MDA is added to each Microelisa stripplate well and incubated. Free components are washed away. The TMB substrate solution is added to each well. Only those wells that contain MDA and HRP conjugated MDA Antibody will appear blue in color and then turn yellow after the addition of the stop solution. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm. The OD value is proportional to the concentration of MDA. You can calculate the concentration of MDA in the samples by comparing the OD of the samples to the standard curve.
Typical Data
The standard curve of QS0067FI is provided for demonstration only. A standard curve should be generated for each set of samples assayed.
Standard Concentration OD Value Average OD Value Blank Well 0ng/ml 0.059 0.0560 Blank Well 0ng/ml 0.053 0.0560 S1 360ng/ml 1.975 1.9695 S1 360ng/ml 1.964 1.9695 S2 240ng/ml 1.388 1.3800 S2 240ng/ml 1.372 1.3800 S3 120ng/ml 0.705 0.6995 S3 120ng/ml 0.694 0.6995 S4 60ng/ml 0.383 0.3800 S4 60ng/ml 0.377 0.3800 S5 30ng/ml 0.197 0.2010 S5 30ng/ml 0.205 0.2010