TEL: +86 571 56623320 EMAIL: [email protected]
DESCRIPTION
CCL11 also named Eotaxin, was initially discovered as an eosinophil-specific chemoattractant. CCL11 plays a central role in both allergic and non-allergic inflammatory reactions by recruiting immune cells such as eosinophils, basophils and Th2 lymphocytes. CCL11 binds to the chemokine receptors CCR2, CCR3 and CCR5, with highest affinity to CCR3. High levels of CCL11 have been described in several chronic inflammatory diseases, such as allergic rhinitis, atopic dermatitis, asthma, gastrointestinal disease and rheumatoid arthritis.
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 232.3 1.9 0.8 2 8 57.0 3.0 5.3 3 8 29.4 0.9 3.1 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 236.4 5.9 2.5 2 16 57.6 3.8 6.6 3 16 30.1 0.9 3.0 RECOVERY
The recovery of mouse/rat CCL11/Eotaxin spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Mouse serum 1:32 124 122-128 1:64 123 120-126 Rat serum 1:80 124 120-127 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The mouse serum was initially diluted 1:4. The rat serum was initially diluted 1:10.)
Mouse serum Rat serum 1:2 Average% of Expected 100 100 Range (%) - - 1:4 Average% of Expected 115 98 Range (%) 112-117 95-102 1:8 Average% of Expected 106 103 Range (%) 100-111 100-106 1:16 Average% of Expected 109 103 Range (%) 104-114 101-106 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human Mouse/Rat CCL11/Eotaxin One-Step ELISA Kit produced at Sunlong Medical™.
Ordering & shipping
- Minimum order: 1 kit
- Lead time: varies with order quantity and specifications. Please consult your local distributor or our sales team for confirmation
- Payment: T/T 100% in advance; PayPal; Alibaba Trade Assurance
- Shipping: FedEx / DHL / UPS express; cold-chain packing available on request
- Price: USD, per kit (48T / 96T options available)
+86 571 56623320
[email protected]
SUNLONG BIOTECH

Scan Wechat Qrcode