TEL: +86 571 56623320    EMAIL: [email protected]

Sunlong Medical™ Mouse E-Cadherin One-Step ELISA Kit
Sunlong Medical™ Mouse E-Cadherin One-Step ELISA Kit
cadherin 1, Cdh1, Ecad, L CAM, Um
Total $498.0
(Vip priceV)
Regular members: $398.4
  • NO.:One-EL286Mo
    Species:Mouse
    Assay range:18.8-1200 pg/mL
    Sensitivity:1.9 pg/mL
    Assay type:Sandwich
    Validity:2-8℃: 6 months; -20℃: 12 months
    NCBI_Gene:12550
    Sample type:Serum, Plasma, Tissue lysate
  • Goods click count:37
  • Product Spec:
  • Quantity: - +
  • Limit points for buying:0 Points
  • Manual
  • Add to cart Inquiry Add to favorite
All categories

Details

Citations

DESCRIPTION

E-cadherin is a transmembrane glycoprotein that connects epithelial cells at adherens junctions. In normal cells, E-cadherin exerts its tumor-suppressing role mainly by sequestering β-catenin from its binding to LEF (Lymphoid enhancer factor)/TCF (T cell factor), which serves the function of transcribing genes of the proliferative Wnt signaling pathway. The loss of E-cadherin expression associated with the epithelial-mesenchymal transition (EMT) occurs frequently during tumor metastasis. Furthermore, Soluble E-cadherin may be biomarkers for the diagnosis and prognosis of cancer.

PRINCIPLE OF THE ASSAY

An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.

PRECISION

Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.

Intra-assay Precision

SamplenMean (pg/mL)SDCV%
18648.223.43.6
28158.12.71.7
3882.81.72.1

Inter-assay Precision

SamplenMean (pg/mL)SDCV%
116656.622.83.5
216160.64.83.0
31681.52.32.8

RECOVERY

The recovery of mouse E-Cadherin spiked to three different levels throughout the range of the assay in various matrices was evaluated.

Sample TypeAverage% of ExpectedRange (%)
Mouse serum1:16,0009286-101
1:32,0009686-107
Tissue lysate1:6408785-89
1:1,2808781-95

LINEARITY

To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The mouse serum was initially diluted 1:2,000. The tissue lysate was initially diluted 1:80.)

Mouse serum(Sample Diluent PT 4B1)Tissue lysate(Sample Diluent PT 4B1-ef)
1:2Average% of Expected100100
Range (%)--
1:4Average% of Expected103107
Range (%)101-10498-116
1:8Average% of Expected110111
Range (%)107-114108-113
1:16Average% of Expected109123
Range (%)100-118120-126

CALIBRATION

This immunoassay is calibrated against a highly purified recombinant mouse E-Cadherin produced at Sunlong Medical™.


Partial purchase records(bought amounts latest0)

No one bought this product
Total 0 records, divided into1 pages First Prev Next Last

User Comment(Total0User Comment Num)

  • No comment
Total 0 records, divided into1 pages First Prev Next Last
Username: Anonymous user
E-mail:
Rank:
Content:
Verification code: captcha

Call us

+86 571 56623320

Address

Room 1-315, Kongle Changqing Building, No. 160 Guangye Road,Gongshu District, Hangzhou City, Zhejiang Province, China

Join Us with

Leave a message
* To protect against spam, please pass the CAPTCHA test below.
© 2005-2026 SUNLONG BIOTECH CO., LTD Copyright, All Rights Reserved.