PD-L1 (programmed cell death ligand 1, also known as CD274 or B7-H1) is a 290 aa type I transmembrane protein. PD-L1 is expressed constitutively on T cells, B cells, DCs, macrophages, mesenchymal stem cells and cultured bone marrow-derived mast cells. In addition, PD-L1 is also expressed on many nonhematopoietic cell types, including vascular endothelial cells, epithelial cells, muscle cells, hepatocytes, pancreatic islet cells, astrocytes in the brain, placental syncytiotrophoblasts, and cells in cornea, iris-ciliary body and retina of eye. PD-L1 is frequently upregulated in a wide variety of solid tumors, including melanoma, ovarian, lung, glioblastoma, breast, and pancreatic cancers. PD-L1 and PD-L2 are two ligands of PD-1. Engagement of PD-1 by PDL1 or PD-L2 transduces a signal that inhibits T-cell proliferation, cytokine production, and cytolytic function. It is critical for the regulation of T cell function during tolerance, autoimmunity and infection. Besides the membrane-bound form, PD-L1 can also exist as a soluble form (sPD-L1) generated either by proteolytic cleavage of membrane-bound form or by translation of alternative spliced mRNA.
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 546.3 13.4 2.5 2 8 127.0 0.5 0.4 3 8 65.2 0.7 1.1 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 569.8 26.7 4.7 2 16 129.2 2.8 2.2 3 16 67.2 2.4 3.6 RECOVERY
The recovery of human PD-L1/CD274 spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human plasma 1:8 102 91-123 1:16 92 79-103 Cell culture supernatant 1:8 118 115-119 1:16 112 100-129 Urine 1:2 103 93-111 1:4 110 108-112 Cell lysate 1:400 108 105-110 1:800 112 108-114 LINEARITY
To assess the linearity of the assay, urine samples were spiked with high concentrations of human PD-L1/CD274 and diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. Human plasma, cell culture supernatant and cell lysate samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The cell culture supernatant was initially diluted 1:2. The cell lysate was initially diluted 1:50.)
Human plasma Cell culturesupernatant Urine Cell lysate 1:2 Average% of Expected 100 100 113 100 Range (%) - - 108-118 - 1:4 Average% of Expected 97 101 117 104 Range (%) 95-99 100-102 116-118 103-105 1:8 Average% of Expected 93 104 126 106 Range (%) 84-102 103-105 124-129 105-107 1:16 Average% of Expected 90 103 126 106 Range (%) 80-101 102-104 121-131 105-106 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human Human PD-L1/CD274 One-Step ELISA Kit produced at Sunlong Medical™.