DESCRIPTION
MMP-8 (Matrix metalloproteinase-8) an enzyme that degrades fibrillar collagens imparting strength to the fetal membranes, is expressed by leukocytes and chorionic cytotrophoblast cells (PMID:15367487). The high-serum MMP-8 levels (>100 ng/mL) had poor cancer-specific survival (PMID: 29808017). MMP8 also has prognosis in patients with gastric cancer (PMID: 30192205).
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 1,282.7 23.8 1.9 2 8 322.9 3.5 1.1 3 8 161.9 3.5 2.2 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 1,314.0 40.3 3.1 2 16 324.5 6.5 2.0 3 16 166.4 6.2 3.7 RECOVERY
The recovery of human MMP-8 spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human plasma 1:8 106 99-111 1:16 117 112-121 Saliva 1:32 106 100-111 1:64 110 104-115 Tissue lysate 1:64 94 88-101 1:128 102 94-115 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The human plasma was initially diluted 1:2. The saliva was initially diluted 1:4. The tissue lysate was initially diluted 1:8.)
Human plasma Saliva Tissue lysate 1:2 Average% of Expected 100 100 100 Range (%) - - - 1:4 Average% of Expected 107 100 103 Range (%) 106-108 98-101 98-108 1:8 Average% of Expected 115 97 109 Range (%) 108-123 95-98 99-119 1:16 Average% of Expected 107 94 115 Range (%) 97-116 92-96 103-127 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human MMP-8 produced at Sunlong Medical™.
Ordering & shipping
- Minimum order: 1 kit
- Lead time: varies with order quantity and specifications. Please consult your local distributor or our sales team for confirmation
- Payment: T/T 100% in advance; PayPal; Alibaba Trade Assurance
- Shipping: FedEx / DHL / UPS express; cold-chain packing available on request
- Price: USD, per kit (48T / 96T options available)
+86 571 56623320