Fibroblast growth factor (FGF)19 is a hormone that regulate metabolic processes particularly during feeding or starvation, thus ultimately influencing energy production. FGF19 is secreted by the intestines during feeding and negatively regulates bile acid synthesis and secretion. FGF19 is regarded as late-acting hormones because its functions are only used after insulin and glucagon have completed their actions. FGF19 shows extensively functional overlap in terms of improving glucose tolerance, insulin sensitivity, weight loss, and lipid, and energy metabolism, particularly in pathological conditions such as diabetes, obesity, metabolic syndrome, and cardiovascular and renal diseases. Most patients with these metabolic diseases exhibit reduced serum FGF19 levels, which might contribute to its etiology.
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 248.6 3.4 1.4 2 8 63.3 1.6 2.5 3 8 31.0 0.9 3.1 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 248.9 3.8 1.5 2 16 63.2 1.6 2.5 3 16 31.3 0.9 2.9 RECOVERY
The recovery of human FGF19 spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human plasma 1:16 92 91-93 Cell culture supernatant 1:160 82 75-87 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The cell culture supernatant was initially diluted 1:40.)
Human plasma Cell culture supernatant 1:2 Average% of Expected 100 100 Range (%) - - 1:4 Average% of Expected 108 96 Range (%) 106-110 88-104 1:8 Average% of Expected 118 98 Range (%) 116-120 87-109 1:16 Average% of Expected 109 92 Range (%) 100-119 90-94 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human FGF19 produced at Sunlong Medical™.