DESCRIPTION
Complement factor D, also known as adipsin, is a highly specific serine protease. Complement factor D is the rate-limiting enzyme in the alternative pathway of complement. It activates C3 convertase by cleaving factor B, which is a key step in the alternative pathway. Complement factor D catalyzes the cleavage of factor B to generate Ba (non-catalytic chain) and Bb (active catalytic subunit), which together with C3(H2O) (or C3b) constitute the active C3 convertase C3(H2O)Bb (or C3bBb).
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 268.6 4.6 1.7 2 8 68.3 2.9 4.2 3 8 32.7 1.6 4.9 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 275.1 9.0 3.3 2 16 68.0 2.8 4.1 3 16 32.2 1.5 4.7 RECOVERY
The recovery of human Complement factor D spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human plasma 1:64,000 89 84-94 Cell culture supernatant 1:256 97 85-112 Urine 1:16 101 88-115 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The human plasma was initially diluted 1:4,000. The cell culture supernatant was initially diluted 1:16. The urine was initially diluted 1:4.)
Human plasma Cell culturesupernatant Urine 1:2 Average% of Expected 100 100 100 Range (%) - - - 1:4 Average% of Expected 98 108 104 Range (%) 97-99 92-123 98-110 1:8 Average% of Expected 109 90 104 Range (%) 101-118 85-94 98-111 1:16 Average% of Expected 91 90 89 Range (%) 80-102 87-93 83-96 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human Complement factor D produced at Sunlong Medical™.