Tissue factor (TF),also named coagulation factor III or F3, is a transmembrane protein that functions as a high-affinity receptor for factor (F)VII and FVIIa. The TF-FVIIa complex is the primary initiator of blood coagulation and plays an essential role in hemostasis. TF is expressed on perivascular cells and epithelial cells at organ and body surfaces where it forms a hemostatic barrier. TF also provides additional hemostatic protection to vital organs, such as the brain, lung, and heart. Under pathological conditions, TF can trigger both arterial and venous thrombosis.
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 258.0 4.3 1.7 2 8 65.0 0.9 1.4 3 8 30.5 1.5 4.9 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 254.5 5.2 2.0 2 16 62.9 2.5 4.0 3 16 29.3 1.8 6.1 RECOVERY
The recovery of human Coagulation Factor III/Tissue Factor spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human plasma 1:4 109 101-115 Urine 1:4 116 112-120 Cell lysate 1:400 109 101-121 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The cell lysate was initially diluted 1:25.)
Human plasma(Sample Diluent PT 3-ef) Urine(Sample Diluent PT 4B1) Cell lysate(Sample Diluent PT 4B1) 1:2 Average% of Expected 100 100 100 Range (%) - - - 1:4 Average% of Expected 116 104 104 Range (%) 104-128 93-115 89-119 1:8 Average% of Expected 115 96 105 Range (%) 110-121 77-115 85-125 1:16 Average% of Expected 120 102 96 Range (%) 120-121 87-116 80-111 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human Coagulation Factor III/Tissue Factor produced at Sunlong Medical™.