TEL: +86 571 56623320 EMAIL: [email protected]
DESCRIPTION
CD80, also known as B7-1, is a type I membrane protein and a member of the immunoglobulin superfamily. It plays a crucial role in T cell activation and regulation by providing a co-stimulatory signal necessary for T cell proliferation and cytokine production. CD80 interacts with CD28 on T cells to promote T cell activation and with CTLA-4 to inhibit T cell activation. This dual role highlights its importance in modulating immune responses. Additionally, CD80 acts as a receptor for adenovirus subgroup B, indicating its involvement in viral infections.
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 2,458.7 72.2 2.9 2 8 583.6 15.1 2.6 3 8 281.4 7.0 2.5 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 2,581.8 141.5 5.5 2 16 612.8 33.9 5.5 3 16 288.4 10.7 3.7 RECOVERY
The recovery of human CD80/B7-1 spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human plasma 1:2 78 71-85 1:4 81 80-83 Cell culture supernatant 1:2 109 107-112 1:4 103 100-106 Urine 1:2 94 87-98 1:4 100 95-105 Cell lysate 1:8 72 71-75 1:16 80 77-82 LINEARITY
To assess the linearity of the assay, cell culture supernatant and urine samples were spiked with high concentrations of human CD80/B7-1 and diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. Human plasma and cell lysate samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay.
Human plasma(Sample Diluent PT4B1) Cell culturesupernatant (SampleDiluent PT 4B1) Urine (SampleDiluent PT 4B1) Cell lysate (SampleDiluent PT 4B1-ac) 1:2 Average% of Expected 100 118 83 100 Range (%) - 112-125 79-87 - 1:4 Average% of Expected 91 113 98 110 Range (%) 86-96 110-115 95-100 109-111 1:8 Average% of Expected 93 104 105 81 Range (%) 86-100 103-104 103-107 79-83 1:16 Average% of Expected 91 92 106 76 Range (%) 83-100 90-93 105-107 75-77 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human Human CD80/B7-1 One-Step ELISA Kit produced at Sunlong Medical™.
Ordering & shipping
- Minimum order: 1 kit
- Lead time: varies with order quantity and specifications. Please consult your local distributor or our sales team for confirmation
- Payment: T/T 100% in advance; PayPal; Alibaba Trade Assurance
- Shipping: FedEx / DHL / UPS express; cold-chain packing available on request
- Price: USD, per kit (48T / 96T options available)