TEL: +86 571 56623320 EMAIL: [email protected]
DESCRIPTION
CD23, also known as low affinity immunoglobulin epsilon Fc receptor, is a transmembrane glycoprotein present on a subpopulation of B lymphocytes in germinal centres, EBV-transformed B-lymphoblastoid cell lines, follicular dendritic cells, and a subpopulation of peripheral blood cells. CD23 has essential roles in the regulation of IgE production and in the differentiation of B-cells. CD23 can be cleaved into soluble fragment (soluble CD23, sCD23) of various sizes displaying pleiotropic biological activities. Increased plasma levels of sCD23 have been reported in allergic disorders, acute viral diseases, lepromatous leprosy, B-CLL, and inflammatory diseases like RA, SLE and sarcoidosis.
PRINCIPLE OF THE ASSAY
An anti-tag antibody is pre-coated onto the bottom of wells. After adding antigen or samples, Tagged-Capture antibody and HRP-conjugated detection antibody, a sandwich complex is formed in the solution. TMB acts as a HRP substrate, and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns the solution yellow. The color intensity is proportional to the quantity of bound protein, which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 8 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 16 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 8 2,326.4 72.7 3.1 2 8 590.9 17.8 3.0 3 8 289.7 9.9 3.4 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 16 2,294.7 80.5 3.5 2 16 568.8 26.8 4.7 3 16 282.2 12.4 4.4 RECOVERY
The recovery of human CD23 spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human plasma 1:4 82 81-83 1:8 89 86-92 Cell culture supernatant 1:8 106 101-112 1:16 98 94-100 LINEARITY
To assess the linearity of the assay, samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay.
Human plasma Cell culture supernatant 1:2 Average% of Expected 100 100 Range (%) - - 1:4 Average% of Expected 111 102 Range (%) 107-115 101-102 1:8 Average% of Expected 107 113 Range (%) 100-114 109-116 1:16 Average% of Expected 98 110 Range (%) 92-105 108-113 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human CD23 produced at Sunlong Medical™.
+86 571 56623320
[email protected]
SUNLONG BIOTECH

Scan Wechat Qrcode