TEL: +86 571 56623320 EMAIL: [email protected]
DESCRIPTION
uPAR/PLAUR is a highly glycosylated GPI-anchored membrane protein. In addition to the membrane-anchored form, uPAR is released from the plasma membrane by cleavage of the GPI anchor and can be found as a soluble form (suPAR). uPAR contains three homologous domains (D1-D3) of which the N-terminal one (D1) represents the uPA-binding domain. After binding to uPAR, uPA cleaves plasminogen, generating the active protease plasmin which is involved in a wide variety of physiologic and pathologic processes. In addition to regulating proteolysis, uPAR has important function in cell adhesion, migration and proliferation. Studies reveal that uPAR expression is elevated during inflammation and tissue remodelling and in many human cancers, in which it frequently indicates poor prognosis. suPAR has been detected in plasma, and increased plasma concentrations of suPAR have been found in patients with some advanced cancers.
PRINCIPLE OF THE ASSAY
Sandwich ELISA structure (Detection antibody labeled with biotin) A capture antibody is pre-coated onto the bottom of wells which binds to analyte of interest. A detection antibody labeled with biotin also binds to the analyte. Streptavidin-HRP binds to the biotin. TMB acts as the HRP substrate and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns solution yellow. The color intensity is proportional to the quantity of bound protein which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 20 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 24 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 20 256.3 12.5 4.9 2 20 512.2 16.3 3.2 3 20 2,009.1 49.8 2.5 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 24 259.7 21.6 8.3 2 24 501.1 28.5 5.7 3 24 1,863.5 154.6 8.3 RECOVERY
The recovery of human uPAR/PLAUR spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human serum 1:4 85 76-90 1:8 93 80-112 Cell culture supernatant 1:16 109 92-119 1:32 100 82-113 Urine 1:16 96 84-111 1:32 84 77-97 LINEARITY
To assess the linearity of the assay, three samples were diluted with Sample Diluent to produce samples with values within the dynamic range of the assay.
Sample Type Range (%) Average% of Expected Human serum 1:4 - 100 1:8 100-102 101 1:16 103-109 106 1:32 111-120 116 Cell culture supernatant 1:8 - 100 1:16 108-112 110 1:32 108-118 113 1:64 113-120 117 Urine 1:8 - 100 1:16 100-109 106 1:32 108-115 111 1:64 111-116 113 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human Human uPAR/PLAUR Sandwich ELISA Kit produced at Sunlong Medical™.
+86 571 56623320
[email protected]
SUNLONG BIOTECH

Scan Wechat Qrcode