TEL: +86 571 56623320 EMAIL: [email protected]
DESCRIPTION
SELP, also named as P-selectin, GMP-140 or CD62P, is a transmembrane glycoprotein that mediates the interaction of activated endothelial cells or platelets with leukocytes. It is an adhesion molecule involved in the pathogenesis of inflammation, thrombosis, and oncogenesis. P-selectin is stored in the alpha-granules of platelets and Weibel-Palade bodies of endothelial cells. Upon cell activation by agonists, P-selectin is transported rapidly to the cell surface. Soluble P-selectin is produced by platelets and endothelial cells and circulates in normal plasma.
PRINCIPLE OF THE ASSAY
Sandwich ELISA structure (Detection antibody labeled with biotin) A capture antibody is pre-coated onto the bottom of wells which binds to analyte of interest. A detection antibody labeled with biotin also binds to the analyte. Streptavidin-HRP binds to the biotin. TMB acts as the HRP substrate and the solution color will change from colorless to blue. A stop solution containing sulfuric acid turns solution yellow. The color intensity is proportional to the quantity of bound protein which is measurable at 450 nm with the correction wavelength set at 630 nm.
PRECISION
Intra-assay Precision (Precision within an assay) Three samples of known concentration were tested 20 times on one plate to assess intra-assay precision. Inter-assay Precision (Precision between assays) Three samples of known concentration were tested in 24 separate assays to assess inter-assay precision.
Intra-assay Precision
Sample n Mean (pg/mL) SD CV% 1 20 1,878.5 46.1 2.5 2 20 483.9 14.9 3.1 3 20 118.6 6.2 5.2 Inter-assay Precision
Sample n Mean (pg/mL) SD CV% 1 24 1,983.3 75.9 3.8 2 24 479.6 18.6 3.9 3 24 104.6 6.4 6.1 RECOVERY
The recovery of human P-selectin spiked to three different levels throughout the range of the assay in various matrices was evaluated.
Sample Type Average% of Expected Range (%) Human serum 1:600 95 78-107 1:1,200 99 87-102 Human plasma 1:100 97 85-104 1:200 104 101-109 Cell culture supernatant 1:2 95 84-113 1:4 94 86-108 LINEARITY
To assess the linearity of the assay, human plasma and serum samples were diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. Cell culture supernatant was spiked with high concentrations of human P-selectin in various matrices and diluted with the appropriate Sample Diluent to produce samples with values within the dynamic range of the assay. (The human plasma samples were initially diluted 1:50. The human serum samples were initially diluted 1:150)
Human plasma(Sample Diuent PT 4) Human serum(Sample Diuent PT 4) Cell culture supernatant(Sample Diuent PT 4-ef) 1:2 Average% ofExpected 100 100 106 Range (%) - - 98-114 1:4 Average% ofExpected 100 100 108 Range (%) 91-107 92-106 100-117 1:8 Average% ofExpected 103 101 111 Range (%) 97-109 95-113 103-118 1:16 Average% ofExpected 106 103 107 Range (%) 99-117 96-115 98-117 CALIBRATION
This immunoassay is calibrated against a highly purified recombinant human P-selectin produced at Sunlong Medical™.
Ordering & shipping
- Minimum order: 1 kit
- Lead time: varies with order quantity and specifications. Please consult your local distributor or our sales team for confirmation
- Payment: T/T 100% in advance; PayPal; Alibaba Trade Assurance
- Shipping: FedEx / DHL / UPS express; cold-chain packing available on request
- Price: USD, per kit (48T / 96T options available)
+86 571 56623320
[email protected]
SUNLONG BIOTECH

Scan Wechat Qrcode