TEL: +86 571 56623320 EMAIL: [email protected]
Assay Principle
This kit employs double antibody sandwich ELISA technology: Capture Antibody is coated on the microplate to capture RLN-3 from samples and standards. After washing, biotin-labeled Detection Antibody is added and incubated, followed by washing to form a "Capture Antibody-Antigen-Detection Antibody" immune-complex. Subsequently, Streptavidin-Horseradish Peroxidase (SA-HRP) is added and incubated. After incubation and washing, TMB Substrate is added for color development. If the target analyte is present in the sample, a blue color develops. Stop Solution is then added to terminate the reaction. During the assay, unbound components are washed away. The Optical Density (OD) is measured at 450 nm using a microplate reader. The intensity of the color is proportional to the RLN-3 concentration in the sample, and the concentration is calculated by plotting a standard curve.
Precision
Intra-assay and inter-assay Coefficients of Variation (CV) are both <10%.
· Intra-assay Precision: Three known concentration samples were assayed 20 times on one plate. The CV of the concentrations was calculated.
· Inter-assay Precision: Three known concentration samples were assayed in 20 replicates across three different plates. The CV of the concentrations was calculated.
Intra-assay Precision
Inter-assay Precision
Sample
1
2
3
1
2
3
n
20
20
20
20
20
20
Mean (pg/mL)
3.69
6.44
13.35
4.1
5.71
11.63
Standard Deviation
0.03
0.15
0.42
0.07
0.17
0.4
CV (%)
0.81
2.32
3.14
1.70
2.97
3.43
Recovery
Recovery was tested by spiking known concentrations of Human RLN-3 into different sample matrices. The recovery range and average recovery are shown below.
Sample Type
Range (%)
Average Recovery (%)
Serum (n=8)
90-102
96
Plasma (n=8)
92-104
98
Cultured Cells (n=8)
98-115
109