SL2079Hu_1 Caspase-3 ELISA detects active p17/p12
|
| Product | Human Caspase-3 (CASP-3)ELISA Kit (Enzyme activity) |
| Cat. No. | SL2079Hu_1 |
The customer is a PhD student in microbiology and is using the Human Caspase-3 (CASP-3) ELISA Kit (enzyme activity assay), catalog number SL2079Hu_1 (batch number [please fill in batch number]), to detect caspase-3 in human serum samples. The reviewer has requested a written explanation of the following: 1) whether the kit detects only activated (cleaved) caspase-3, only procaspase-3, or total caspase-3 (both forms); 2) what antibody clones or epitopes are used for the capture and HRP-conjugated detection antibodies, and whether they specifically recognize the cleaved fragments (p17/p12); 3) the kit is labeled as enzyme activity, but the instructions describe it as a sandwich ELISA, with 450 nm absorbance proportional to CASP-3 concentration; how is enzyme activity determined, and how is the U/L unit defined and calibrated (reference standard and its source).
1. Detection Target
This kit specifically detects activated, cleaved Caspase-3 (p17/p12 fragments), does not recognize full-length inactive procaspase-3, and does not detect total Caspase-3.
2. Capture Antibody and HRP-labeled Detection Antibody Information
Both the capture antibody used for coating and the HRP-conjugated detection antibody are developed against the p17 large subunit of human activated Caspase-3:
- Capture antibody: The epitope is located in the p17 fragment generated after cleavage, with no cross-reactivity to full-length procaspase-3; the antibody clone number is proprietary corporate information and is not disclosed.
- HRP-labeled detection antibody: Recognizes an epitope on the p17 subunit of activated Caspase-3 and does not bind the precursor domain or full-length procaspase-3.
Neither antibody recognizes uncleaved procaspase-3.
3. Detection Principle and Definition of U/L Unit
Regarding your point that “a sandwich ELISA is labeled with enzyme activity,” the following explanation is provided:
This kit is a sandwich immunoassay, and the A450 absorbance signal is proportional to the activated Caspase-3 protein content in the sample.
The “enzyme activity” in the kit name originates from the calibration method of the standard: the standard supplied with the kit is calibrated using recombinant catalytically active human Caspase-3 (p17/p12 heterodimer); the true catalytic activity of this recombinant active Caspase-3 is first measured using the fluorescent substrate Ac-DEVD-AFC by a biochemical enzyme activity method, and then the active Caspase-3 protein concentration is converted into enzyme activity units U/L.
Unit definition in this system: 1 unit (U) refers to the amount of activated Caspase-3 required to hydrolyze 1 nmol Ac-DEVD-AFC substrate per minute under the specified buffer system at 37 ℃.
Therefore, the U/L value given by this ELISA is an activity-equivalent value calibrated with an activity standard, and is different from traditional enzyme activity assays that directly monitor substrate hydrolysis rate; the kit microplate does not contain chromogenic/fluorescent substrate, and no enzymatic catalytic reaction is performed on the plate.
· Sample type: human serum samples
· Other limitations: This kit specifically detects activated, cleaved Caspase-3 (p17/p12 fragments), does not recognize full-length inactive procaspase-3, and does not detect total Caspase-3; both the capture antibody and the HRP-labeled detection antibody target the p17 large subunit of human activated Caspase-3, and the antibody clone numbers are proprietary corporate information and will not be disclosed; U/L is the activity equivalent value calibrated against an activity standard; the microplate does not contain chromogenic/fluorescent substrates, and no enzymatic catalytic reaction is performed on the plate.